Summary
Type: TLC
Application #: 400330
Catalog#: 1715026 - Silica G HR TLC Plates, glass backed, 250um, 20x20cm, 25/pk
Detection: Spray with 0.5% solution of dichloroquinone chloroimide in ethanol, heat to 105 C (10 min), then treat with ammonia vapour (see chapter visualization reagents)
Conditions: Eluent: petroleum ether (40-60 C) / toluene / glacial acetic acid (2:2:1, v/v)
Developing: 25 min under exclusion of light,
time: chamber saturation
Authors: TLC department
Source: TLC department